ISA-60M Oxygen Monitor
1-3 Channels
with Remote MRI-5175 O2 Sensor/Transmitter

 

Oxygen Monitor Alarm for MRI Installations

 

Maintaining the quality of breathing air in a Magnetic Resonance Imaging (MRI) facility is an important safety consideration. Cryogenic gas used to maintain the magnetic field is highly dangerous if released into an enclosed space. Inert gas can create a hazardous situation by reducing the concentration of oxygen in breathing air to unsafe levels.

The silent, odorless, and invisible hazard of low oxygen concentrations can incapacitate or even kill. Reduced oxygen levels start by causing fatigue, confusion, and disorientation without individuals being aware of their peril. If oxygen levels are further reduced, fatalities may result in minutes.


Enmet spacer

Specifications

GENERAL SPECIFICATIONS:


Temperature: 32° to 104°F (0° to 40°C)

Relative Humidity: 10-99% RH, non-condensing

Pressure, Atmospheric: 20 to 36 inHg (68 to 122 kPa)

Approval: Classified to UL 60601-1
IEC 60601-1 & SCA 22.2, No. 601-1




ISA-60M MONITOR SPECIFICATIONS:


Standard Alarm Points: 19.5% & 17% decreasing oxygen

Display: 2 line, 16 character, dot matrix LCD

Alarms: Visual: LEDs, Audible: piezo electric

Horn: 95 dB at 2 feet

Alarm Relays: 5 programmable oxygen alarm relays, plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps (resistive load only) at 110 VAC.

Analog Output: 4-20 mA

Digital Output: RS-232 Modbus and RS-485 Modbus

Connections: Strain relief (2 supplied) for 0.20" to 0.35" OD Cable

Operating Power: 100 to 240 VAC and/or 12 VDC, 15 Watts

Enclosure: Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X

Size: 10.5 H x 8.5 W x 5.8 D inches (27 x 22 x 15 cm)

Weight: 8 lbs. (3.6 kg)

Channel Capacity: From 1 to 3 MRI-5175 Remote Oxygen Sensor/Transmitters can be connected to one ISA-60M monitor utilizing 4-20 mA singnals. Up to four MRI-5175 can be displayed at the ISA-60M using RS-485 Modbus.

MRI-TRANSMITTER:


Visual Gas Alarm: 3 Red LEDs

Visual Power/Fault: Green/Red LED

Analog Output: 4-20 mA

Digital Output: RS-485 Modbus

Power: 24 VDC from ISA-60M

Optional Audible Alarm: 105 dB at 12 inches (30 cm)

Connection to ISA-60M: Strain relief (1 supplied) for 24 VDC, 4-20 mA, 0.20" to 0.35" OD Cable

Enclosure: Fiberglass-reinforced polyester non-magnetic hardware

Dimensions: 5 x 5 x 4 inches (2 x 2 x 1.6 cm)

Weight: 2 lbs. (0.9 kg)




OXYGEN SENSOR, SUPPLIED WITH MRI-5175:


Type:
Electrochemical

Resolution: 0.1% by Volume Oxygen

Range: 0 to 30.0% by Volume Oxygen

Typical Life: 4 to 5 years




SENSOR CABLE:


Cable, MRI-5175 to ISA-60M:
3-wire, shielded cable, typically 18 to 20 AWG, user supplied, UL Listed type CM, PLTC or TC



For detailed information including drawings with external and internal features, dimensions and ordering information see product literature: ISA-60M Literature

Enmet spacer

Standard Oxygen Monitors

The Model ISA-60M is specifically designed for use at Medical MRI Installations. For a basic oxygen monitor, see our Model AM-5175 single-point oxygen monitor.

back to top