Calibration Equipment, Replacement Parts and Accessories are available. Call us for More Information.
GSM-60 Gas Sampling Monitor with Internal Pump and Sensors
Sampling Monitor, with Internal Pump and up to 4 Sensors
Also Accepts 4-20 mA Input from Remote Sensor/Transmitter
Designed for Monitoring Ductwork, Tank Headspaces, Scrubber Exhausts, etc.
Many Instrument Configuration Options Available for VOCs, Oxygen, H2, CO2, CO and many Other Toxic Gases
Internal Pump and Sensors for Monitoring Scrubber Exhaust, Tank Headspaces, Ductwork, or other Enclosed Spaces, plus the Capability of Ambient Air Monitoring with Remote Sensor/Transmitters
Monitors Up to 4 Gases with a Combination of Internal Sensors and Remote Sensor/Transmitters
Designed for use in Industrial, Aerospace, Medical, Pharmaceutical,
Semiconductor, and General Process Applications
Programmable Relay Contacts
Adjustable Alarm Points
Large Easy-to-Read Display
RS-232/RS-485 Modbus Communication
4-20 mA Inputs and Outputs
Can be Configured on Request for Reactive Gases, including O3, HF, Cl2, etc.
The GSM-60 is a versatile instrument, which, in addition to having an internal sampling pump and sensors, can also accept inputs from remote gas sensor/transmitters. The system can be custom configured to monitor a variety of conditions, including VOCs, CO, CO2, oxygen, as well as many toxic and other gases. The instrument has a user-friendly interface for all maintenance and operation functions, and it is protected by a compact and durable enclosure for process environments. Applications include: medical, pharmaceutical, aerospace and process manufacturing industries in general.
General Specifications
Display:
2 line, 16 character, dot matrix LCD
Alarms:
Visual: LEDs, Audible: piezo electric
Horn:
95 dB at 2 feet
Alarm Relays:
5 programmable gas relays plus fault. All relays are programmable latching or non-latching, dry SPDT, 10 amps, (resistive load only) at 110 VAC.
Operating Power :
100 to 240 VAC and/or 12 VDC, 15 Watts
Flow Rate:
1 Lpm (pump/internal sensors)
Sample Inlet Connection:
Female quick release, supplied with male quick release for 1/4" I.D. tubing. Tubing available from ENMET on request
Enclosure :
Thermoplastic box with clear, hinged from cover, designed for NEMA 12 and 4X
Size:
10.5 x 8.5 x 7.8 inches
Weight:
9 lbs.
For more detailed information including instrument drawing with interior features see product literature: GSM-60 Literature
Sensor Specifications
Gas
Typical Range (1)
Display Resolution
Example (2)
Low, High (3)
Alarm Point LEDs
VOCs
0-100 ppm
0.1 ppm
10, 20 ppm
Dew Point
-112° to 68°F
1°F
-40°, +39°F
N2O
0-2000 ppm
20 ppm
100, 500 ppm
HC
0-100% LEL
1% LEL
10, 20% LEL
HC/VOCs
0-500 ppm
1 ppm
100, 200 ppm
CO2
0-5000 ppm
20 ppm
1000, 2000 ppm
O2
0-30% by Vol
0.1% by Vol
19.5%, 23.5% by Vol
CO
0-500
1 ppm
50, 200 ppm
NH3
0-100
1 ppm
25, 75 ppm
SO2
0-30 ppm
0.1 ppm
2, 10 ppm
H2
0-2000 ppm
1 ppm
200, 1000 ppm
H2S
0-100 ppm
1 ppm
10, 50 ppm
NO
0-100 ppm
1 ppm
25, 75 ppm
NO2
0-30 ppm
0.1 ppm
3, 10 ppm
ETO
0-10 ppm
0.1 ppm
3, 9 ppm
NOTES: (1) Examples of typical ranges. Other ranges may be available on request.
(2) Examples of typical alarm points. Other alarm points available on request.
(3) High and Low alarm points are user programmable.
For detailed information including drawings with external and internal features, dimensions and ordering information see product literature: GSM-60 Literature